Troubleshooting guide is divided according to the main themes. This division will allow for electronization and implementation into the GeneProof Web site, including a continuous update. Information sources for the updates are OTRS, discussion with customers and others. Troubleshooting data are updated repeatedly.
The troubleshooting logic copy the diagnostics process in standard molecular diagnostic laboratories. The general topic is always written in capital letters, which is then further expand according to the specific problem.
| Possible Cause | Solution |
|---|---|
| Degraded MasterMix | Store at -20°C, thaw at 2 - 8°C until fully thawed. Vortex and short spin before use. |
| Unverified cycler | Use verified cycler. List of verified cyclers here. |
| Incorrect consumables | Use recommended consumables. List of consumables in device manual. |
| Clear PCR strips on BioRad CFX96 | Use white PCR strips on BioRad CFX96 |
| Incorrect transport medium | Use TE buffer or Nuclease-free water. |
| Air bubbles in PCR mix solution | Spin PCR strip before PCR. |
| Photosensitive MasterMix | Minimize exposure of the MasterMix to light during handling to avoid loss of fluorescent signal intensity. |
|
Combination of parameters with unequal fluorescence on Rotor-Gene Q |
Place the positive control from the problematic PCR kit in well 1 and enable Auto-Gain Optimization, or run the problematic kit in a separate experiment. |
| Possible cause | Solution |
|---|---|
| Incorrect PCR setup | Contact support. Templates can be provided. |
| Incorrect consumables | Use recommended consumables. List of consumables in device manual. |
| Incorrect transport medium | Use TE buffer or Nuclease-free water. |
| Photosensitive MasterMix | Minimize exposure of the MasterMix to light during handling to avoid loss of fluorescent signal intensity. |
| Degraded PCR reagents | Transport the PCR kit on dry ice. |
|
Combination of parameters with unequal fluorescence on Rotor-Gene Q |
Place the positive control from the problematic PCR kit in well 1 and enable Auto-Gain Optimization, or run the problematic kit in a separate experiment. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Insufficient extraction | Use validated extraction. |
| Unvalidated specimen | Check list of validated specimen in IFU. |
| Mishandling with sample | Check sampling and sample storage in IFU. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Invalid curves caused by software | Contact support. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Contamination | Clean workspace. |
| Cross-talk | Contact support. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Incorrect volume | Add 10 % of elution volume to sample before extraction. |
| Degraded IC | Store at -20 °C, thaw at 2 - 8 °C until fully thawed. |
| Insufficient extraction | Use validated extraction. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible cause | Solution |
|---|---|
| Possible competition | Normal - still valid result. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible cause | Solution |
|---|---|
| Failed extraction | Use validated extraction. Repeat whole analysis. |
| PCR inhibition | Repeat analysis of failed samples. |
| PCR inhibition | When working with GeneProof PathogenFree RNA Isolation Kit, eluate in 100 µl to dilute inhibitors. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Contamination | Clean workspace. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Degraded Positive Control | Store at -20 °C, thaw at 2 - 8 °C no longer than 30 min. Vortex and short spin before use. |
| Degraded MasterMix | Store at -20 °C, thaw at 2 - 8 °C no longer than 30 min. Vortex and short spin before use. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Cross-talk | Set threshold (TH) above curve in HEX. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Incorrect pipetting | Repeat PCR. Valid calibration curve values - R^2 ≥ 0,98; SLOPE between -3,8 to -3,1. |
Was the answer helpful?
YES NO Contact our technical support.
| Possible Cause | Solution |
|---|---|
| Incorrectly set threshold | Set threshold lower/set threshold just above negative control |
Was the answer helpful?